Application |
Flexible and intuitive chromatography system for fast purification of proteins, peptides, and nucleic acids from microgram to gram levels of target product. |
Application supported |
Affinity chromatography, size exclusion chromatography (SEC, also known as gel filtration), ion exchange chromatography, hydrophobic interaction chromatography, and reversed phase chromatography (RPC). For RPC runs use only Frac F9-R. |
Flow rate |
0.001 to 25 mL/min (normal range), 0.01 to 50 mL/min (column packing flow) |
Operating pressure max. |
20 MPa (2900 psi) |
Operating pressure (sample pump) max. |
N/A |
Tubing i.d. (flow path) |
0.5 mm |
Multiple wavelength detection |
Up to 3 wavelengths |
UV wavelength |
190 to 700 nm |
Flow cell path length |
2 mm (optional 0.5 mm and 10 mm) |
pH monitoring |
0 to 14 (optional) |
Multiple sample injection |
Upgradeable to 5 samples |
Autosampler injection |
No |
Automatic buffer preparation |
No |
Buffer selection |
Upgradeable to 14 inlets |
pH stability, cleaning-in-place (CIP) |
0 to 14 |
pH stability, operational |
2 to 12 |
Multistep purification |
Yes, (optional) |
Relative humidity |
20% to 95% (noncondensing) |
Viscosity |
0.35 to 10.0 cP (system pump) |